BATTERY HANDLING AND USAGE
When the monitor is connected to the "iHealth MyVitals" APP, the battery charge will be displayed in the APP. If the power is less than 25%, please charge the battery.
The monitor will not work until the battery has enough power.
- When the monitor needs charging, please connect the monitor to a power source.
- You should charge the battery when the battery is less than 25% charged. Overcharging the
battery may reduce its lifetime. - When in charging mode, the charging status will be displayed on the LED screen. See the table
below for details
Monitor Status | Status Indicator |
Charging | battery /charge symbol flashing slowly |
Fully charged | battery/charge symbol steady |
Battery charge <25% | battery /charge symbol flashing fast |
Battery low | battery/charge symbol flashing fast |
Troubleshooting
Problem | Possible Cause | Solution |
Low Battery | Battery do not have enough power | Charge the battery |
LED display reads "ErO" | Pressure system is unstable before measurement | Retest, make sure not to move your arm or the monitor |
LED display reads "Er1" | Failed to detect systolic pressure | |
LED display reads "Er2" | Failed to detect diastolic pressure | |
LED display reads "Er3" | Pneumatic system blocked or cuff is too tight during inflation | Apply the cuff correctly and try again. |
LED display reads "Er4" | Pneumatic system leakage or cuff is too loose during inflation | |
LED display reads "Er5" | Cuff pressure above 300mmHg | Measure again after five minutes. If the monitor is still abnormal, please contact customer support. |
LED display reads "Er6" | More than 160 seconds with cuff pressure above 15 mmHg | |
LED display reads "Er7" | memory accessing error | |
LED display reads "Er8" | Device parameter checking error | |
LED display reads "ErA" | Pressure sensor parameter error | |
LED display reads " |
Bluetooth communicate error | Reset monitor by pressing the START button and holding for about 10 seconds, then connect the mobile device correctly And try again, If the monitor is still abnormal, please contact customer support. |
LED display reads an abnormal result | The cuff position was not correct or it was not properly tightened | Review the cuff application instructions end retest |
Body posture was not correct during testing | Review body posture instructions and retest | |
Speaking, moving arm or body, being angry, excited or nervous during test | Retest when calm; avoid speaking or movement during the test | |
Bluetooth connection unstable | Bluetooth connection unsuccessful, monitor is abnormal, or strong electromagnetic interference is present | Resel iOS/Android device. Reset monitor by pressing the START/STOP button and holding for about 10 seconds. Make sure the monitor and iOS/Android device are away from other electrical equipment Please see GENERAL SAFETY AND PRECAUTIONS in the manual |
No response when you press button | Incorrect operation or strong electromagnetic interference | Press the START/STOP button and hold for about 10 seconds to reset the device. |
CARE AND MAINTENANCE
- If this monitor is stored near freezing temperatures, allow it to return to room temperature before use.
- If the monitor is not used for a long time, please be sure to fully charge it every month.
- No monitor component needs to be maintained by the user. The circuit diagrams, component part lists, descriptions, calibration instructions, or other information which will assist the user's appropriately qualified technical personnel to repair those parts of the equipment which are designated for repair can be supplied.
- Clean the monitor with a dry, soft cloth or a moistened and well wrung soft cloth using water, diluted disinfectant alcohol, or diluted detergent.
- The monitor can maintain the safety and performance characteristics for a minimum of 10,000measurements or three years of usage, and the cuff integrity is maintained after 1,000 open-close cycles.
- The battery can maintain the performance characteristics for a minimum of 300 charge cycles.
- It is recommended that if the cuff is used in a hospital or a clinic, it be disinfected twice a week. Wipe the inner side (the side that contacts skin) of the cuff with a soft cloth lightly moistened with Ethyl alcohol (75-90%). Then air-dry the cuff.
- It is recommended that product performance be checked every 2 years or after each repair. Please contact the customer service.
- The monitor requires 6 hours to warm from the minimum storage temperature between uses until the monitor is ready for its INTENDED USE when the ambient temperature is 20 °C.
- The monitor requires 6 hours to cool from the maximum storage temperature between uses until the monitor is ready for its INTENDED USE when the ambient temperature is 20 °C.
- Not servicing/maintenance while the monitor is in use.
Do not drop this monitor or subject it to strong impact.
Avoid high temperature and direct sunlight. Do not immerse the monitor in water as this will result in damage to the monitor.
Do not attempt to disassemble this monitor.
Battery replacement should only be performed by a qualified iHealth technician. To do otherwise will void your warranty and possibly damage your unit.
Cuff replacement should only be performed by a qualified iHealth technician. To do otherwise will possibly damage your unit.