Connectivity: Bluetooth BLE 4.1
Screen: LED
Dimensions: 97 mm x 35 mm x 20 mm
Weight: 32.5 g (battery included)
Measurement method: amperometric technology that uses glucose oxidase
Power: 3.7 VDC
Measuring range: 20-600 mg/dL (1.1-33.3 mmol/L)
Accuracy:The blood glucose monitoring system meets the accuracy requirements of standard EN ISO 15197:2015 which states that 95% of the readings less than 100 mg/dL must be within ± 15 mg/dL of the reference value and 95% of the values greater than 100 mg/dL must be within ± 15% of the reference value.
Blood source: fresh capillary whole blood
Blood Volume: 0.7µL minimum
Conditions for use:
- Temperature: 10 to 40°C
- Humidity: 25% - 85%
Storage and transportation conditions (glucometer):
- Temperature: -20°C to 55°C
- Humidity: < 80 % RH
Storage and transportation conditions (EGS-2003 test strip):
- Temperature: 4 to 30°C
- Humidity: < 80 % RH
Lifespan: Five Years
How many measurements can I make on a full charge?
Abour 200